EVOLUTION GC-MS/MS SPECIFICATIONS
CHROMSYS LLC
Dear GC-MS user,

Following a recent upgrade to the Evolution Triple Quadrupole system hardware, we are happy to announce below the latest specifications of our Evolution Triple Quadrupole upgrade for Agilent 5975 systems. If you're interested in the system's performance for your analyte of interest, feel free to send us a sample and concentrated standard, if possible. (Contact us to arrange this). 
 
Evo smallBasic specifications (more detailed specifications on the Evolution Data Sheet) for a new upgraded Agilent 5975 
 
Mode:
EI (PCI/NCI Optional)
Resolution: Unit mass adjustable by tune, 0.5 to 3 Da
Scan rate (electronic): 6,250 u/s
SRM(MRM) speed: 200 transitions/250 ms
Minimum SRM (MRM) dwell: 1 ms
Maximum mass: 800 u
EI MS/MS sensitivity: Injection of 100 fg of octafluoronaphthalene (OFN) will produce a >100:1 RMS S/N for the transition of m/z 272 to the fragment ion at m/z 222 using Autotune parameters. (Diffusion pump systems: 1pg OFN will produce a >500:1 RMS S/N).
EI Scan sensitivity: 1 pg OFN scanning from 90-300 u will give at
nominal m/z 272 ion >300:1 S/N. (Diffusion pump systems: 1pg OFN will produce a >200:1 S/N).
PCI Scan sensitivity: 100 pg BZP will give at nominal m/z 183 ion >100:1 S/N (using methane) N/A for Diffusion pump systems.
NCI Scan sensitivity: 200 fg OFN will give at nominal m/z 272 ion >500:1 S/N (using methane) N/A for Diffusion pump systems.
Mass filters: Q1: Agilent proprietary monolithicInside evo
hyperbolic gold-coated quadrupole,
Q3: Ultra-high precision quadrupole
Collision cell: 90° square quadrupole patented IonRail; low pressure design.
Collision cell gas: Argon or nitrogen
Collision energy: Up to 65 eV
Detector: Triple-Axis HED-EM with extended-life EM
Tuning: Autotune, Quicktune or Manual
Pumping system: 65 L/s diffusion pump, 70 L/s or 262 L/s turbomolecular pump with 2.5 m3/h mechanical pump.
Acquisition control: Agilent Chemstation software with Evo Driver
Data analysis and reporting: Agilent Chemstation software with Evo Driver
Dimensions: 30 cm (w) x 71,5 cm (d) x 41 cm (height, front); 80 cm (height, back). Additional space for data system and printer.
Weight: 70 kg/154 lbs
 
Price for a full system (including 7890 GC) starts at $129,000. Upgrade only- $79,000. Discounts available for academic institutions.
If you have any questions about the specifications above, or if you would like a quote, please give us a call or email us at info@chromsys.com
 
Sincerely,
 
Ingo Christ
General Manager
Agilent Channel Partner img